Technical Updates on UTEVA, TEVA and TRU Resins Eichrom Technologies, Inc.

Download Report

Transcript Technical Updates on UTEVA, TEVA and TRU Resins Eichrom Technologies, Inc.

Technical Updates on UTEVA, TEVA and TRU Resins

Anil Thakkar & Michael Fern Eichrom Technologies, Inc.

Paris May 14, 2002

Topics

• UTEVA Resin: – Decontamination of Po-210 – U-232 yields • TEVA Resin: – Decontamination of U-nat/Th in Tc-99 region • TRU Resin: – Am recoveries

U-232 Yields

• Unrealistic high U-232 yields reported with UTEVA resin in samples with high in Pb-210

Current Procedure

1) Load sample on UTEVA resin with 10 ml of 3M HNO 3 /1M Al(NO 3 ) 3. with 2 mL of 0.6M Ferrous sulfamate and 200 mg of ascorbic acid .

2) Rinse sample beaker with 5 mL of 3M HNO 3 3) Rinse column with 5 mL of 3M HNO 3 4) Rinse column with 5 mL of 9M HCl 5) Rinse column with 20 mL of 5M HCl/0.05M Oxalic 6) Elute U with 15 mL of 0.01M HCl

Effect of Ferrous

210 Pb/ 210 Po spike (1100 pCi) Method Ferrous Activity in U Region (dpm) 230 No ferrous 3.5

Effect of Additional 3M HNO 3 210 Pb/ 210 Po spike Rinse Additional Rinse (mL) 0 10 20 30 Activity in Uranium Region (dpm) 92 pCi spike no ferrous 1080 pCi spike ferrous 3.0

230 1.6

- 0.23

- - 28

Effect of Uranium Strip Solution no ferrous added 210 Pb/ 210 Po spike (1100 pCi) Uranium Strip Solution 0.01M HCl 0.1M HCl 1M HCl Activity in Uranium Region (dpm) No ferrous 3.5

ferrous 230 1.3

0.16

- 1.84

Additional Rinse/1M HCl Strip Combo Method Original procedure 30 mL HNO 3 Rinse only 1M HCl Strip only 30mL rinse + 1M Strip Activity in Uranium Region (dpm) 230 28 1.84

Revised Procedure

1) Load sample on UTEVA with 10 ml of 3M HNO 3 /1M Al(N0 3 ) 3 with 2 mL of 0.6M Ferrous sulfamate and 200 mg of ascorbic acid 2) Rinse sample beaker with 5 mL of 3M HNO 3 3) Rinse column with 5 mL of 3M HNO 3 4) Rinse column with 10 mL of 3M HNO 3 (3x) 5) Rinse column with 5 mL of 9M HCl 6) Rinse column with 20 mL of 5M HCl/0.05M Oxalic 7) Elute U with 15 mL of 1M HCl

Summary

A good decontamination of Po-210 in U-232 region can be achieved by additional 3M HNO 3 rinse on UTEVA and eluting uranium selectively with 1M HCl

TEVA Column: Tc-99 Issues

• Problem reported by Darrin Mann (BWXT, Oak Ridge, Tn) with bias high Tc-99 recoveries in samples with natural uranium.

• Th-234 interference was detected in Tc-99 region.

BWXT’s Tc-99 Procedure

• Solid samples are digested and dissolved finally in 20 ml of 0.1M HNO 3 • Load sample on TEVA column • Rinse column with 25 ml of 0.1M HNO3 • Extrude resin from the column in a LSC vial.

• Add cocktail and count on LSC.

BWXT’s Procedure tested at Eichrom

• Spiked 20 ml of 0.01M HNO 3 U-234/238 with ~2000 dpm of • Load on TEVA Resin column • Rinse with 25 ml of 0.1M HNO 3 • Extrude resin in LSC vial • Added 15 ml of cocktail and count

Possible Solutions for Decontamination

• Use of fluoride to complex the residual Th-234 on the column • Two fluoride solutions tested: – Hydrofluoric acid – Sodium fluoride

Decontamination of Th-234

• Spiked 20 ml of 0.01HNO

3 with ~2000 dpm of U-234/238 • Load on TEVA Resin column • Rinse with 5 ml of 0.1M HNO 3 • Rinse with 25 ml of 0.5M HF/0.02M HNO 3

OR

with 25 ml of 1M NaF/0.02 HNO 3 • Rinse with 5 ml of 0.1M HNO 3 • Extrude resin in LSC vial • Added 15 ml of cocktail and count

BWXT’s ProcedureTest Results Rinse volume used=25 mL TEVA Lot # 0.1M HNO 3 (CPM) TEA06309 72 0.5M HF /0.02M HNO 3 (CPM) 14.84

18.9

13.1

1M NaF /0.02M HNO 3 (CPM) Not tested 15.36

TE0K TEVA Cl form Blank 24.5

14.7 + 0.9

15.36

14.7 + 0.9

15.0

14.7 + 0.9

Conclusion

• Decontamination of Th-234 in Tc-99 region can be accomplished with additional fluoride rinse on the TEVA Resin columns • Fluoride in the form of 1M NaF or 0.5M HF solutions could be used to remove the residual amounts of Th-234

TRU Resin: History • Historical problems with Am recoveries from TRU Resin (HCl strip.) • Suspected Extractant Leaching handled with wet ashing of HCl Strip Solution

TRU Resin Cartridges

1) TRU Resin 2) TRU Resin with a ‘post’ Prefilter layer

Am-243 Spike Test

1) Spike 10 ml of 3M HNO 3 with 6.52 dpm of Am-243 2) Load solution on TRU cartridges (assisted with VBS, flow rate 1ml/min) 3) Rinse with 10 ml of 3M HNO 3 4) Rinse with 5ml of 0.5M HNO 3 5) Strip Am with 9M + 4M HCl

Cerium Fluoride Precipitation

Four different preparation prior to CeF 3 ppt.

– 1) Digested with Aqua-regia – 2) Undigested – 3) Changed molarity (~0.5 M HCl) – 4) Digested with 10% H 2 SO 4

Am-243 recoveries (n=4)

Resin Type

TRU

Am Strip Solution

9M HCl + 4M HCl

Digested with Aqua regia

89 % + 3%

No digestion

48 % + 3%

Changed molarity

97 % + 7%

Digested with H 2 SO 4

84 % + 6% TRU with prefilter layer 9M HCl + 4M HCl 89 % + 5% 62 % + 5% 97 % + 3% 82 % + 23%

RETEST: Am-243 recoveries (n=4)

Resin Type

TRU

Am Strip Solution

9M HCl + 4M HCl TRU with prefilter layer 9M HCl + 4M HCl

Changed molarity First Run

97 % + 7%

Changed molarity RERUN

93 % + 10% 97 % + 3% 95 % + 4%

ACW03 tested for Am

• 0.5 L of Eichrom tap water spiked with Am-243 • Precipitated Am with calcium phosphate • Scavenge ppt. and dissolve in the load solution • Followed ACW03 procedure using UTEVA/TRU cartridges • Collect Am fraction

ACW03…continued

6) Evaporate Am fraction to dryness 7) Add 0.5 ml of conc. HCl and dilute to 15 ml of DI water 8) Perform CeF 3 ppt.

Tap Water: Am-243 recoveries (n=5)

Resin Type

UTEVA /

TRU Am Strip Solution

9M HCl + 4M HCl UTEVA /

TRU

with prefilter layer 9M HCl + 4M HCl

Changed molarity

89 % + 10% 89 % + 11%

Conclusion

• Am recoveries improved by changing the concentration on HCl prior to CeF 3 • TRU resin cartridge performed well once the HCl concentration was changed